Gasmet On-line Series includes industrial multicomponent gas analyzers for continuous monitoring applications. The Gasmet CX4000 incorporates a Fourier transform infrared spectrometer, a temperature controlled sample cell, and signal processing electronics. The analyzer is fully equipped for fixed installations and it offers versatility and high performance for all industrial users.
Gasme在线测试系列包括工业多组分气体分析仪应用于连续在线监测领域。Gasmet CX4000 由傅里叶变换红外光谱仪、 样品温度控制单元和信号处理电子系统组成。该分析仪为所有工业用户提供多功能和高性能的成套装备。
The Gasmet CX4000 is designed for continuous emission monitoring. It is an ideal tool to analyze trace concentrations of pollutants in wet, corrosive gas mixtures. The sample cell can be heated up to 180 °C. Sample cell absorption path length is selected according to the application.
shlf1314Gasmet CX400分析仪是专门为检测烟气的连续排放而设计的。它是来分析潮湿的腐蚀性气体混合物中痕量污染物浓度的理想工具。样品池可以被加热到180 ℃,吸收光谱样品单元可根据客户实际应用调整选择。
The Gasmet CX4000 allows simple calibration using only single component calibration gases. The user can easily configure the analyzer for a new set of compounds.
Gasmet CX4000分析仪允许使用只组分标准气体标定。使用者很容易配置新的测试气体组分。
General parameters 常规参数
Measuring principle:shlf1314Fourier transform infrared, FTIR
shlf1314Performance:Simultaneous analysis of up to 50
gas compounds
Response time, T90:shlf1314Typically < 120 s, depending on
the gas flow and measurement
Time
shlf1314Operating temperature:5 - 30°C, non-condensing
air conditioning recommended
Storage temperature:-20 - 60°C, non-condensing
Power supply:100-115 or 230 V / 50 -60 Hz
Power consumption:300 W max, continuous 150
shlf1314测量原理:FT-IR(傅里叶变换红外光谱分析)
shlf1314性能:可同时测量多达50种气体组分
shlf1314响应时间T90:< 120 s,取决于气体流速和测量时间
shlf1314工作温度:5 - 30°C,无冷凝,建议使用空调以保持恒温。
shlf1314贮存温度:-20 - 60°C,无冷凝
shlf1314电源:100-115v or 230 V / 50 -60 Hz
功率:最大功率为300W,连续工作功率为150W
Spectrometer 分光仪
Resolution:4 cm-1
Scan frequency:10 scans / s
Detector:Peltier cooled MCT
Source:SiC, 1550 K
Beamsplitter:ZnSe
Window material:ZnSe
Wave number range:500 - 4 200 cm-1
分辨率:4 cm-1
shlf1314扫描频率:10 次/s
检测器:Peltier 制冷 MCT
红外光源:SiC, 1550 K
分束器:ZnSe
窗口材质:ZnSe
波数范围:500 - 4200 cm-1
Sample cell 样品室
Structure:Multi-pass, fixed path length 5.0 m
Standard material:100 % rhodium coated aluminium
Mirrors:Fixed, protected gold coating
Volume:0.4 liters
Connectors:Inlet Swagelok 6 mm
shlf1314Outlet Swagelok 8 mm
Gaskets:Viton® O-rings
shlf1314Temperature:180 °C, maximum
Window material:BaF2
shlf1314结构:多册反射光程,固定路径长度为5.0m
标准材质:100%铑镀铝
反射镜:固定,黄金镀层保护
容积:0.4L
shlf1314连接器:入口接头套管6mm,出口接头套管8mm;
垫片:O型氟化橡胶
温度:最高180℃
窗口材质:BaF2
shlf1314Measuring parameters 测量参数
Zero point calibration:24 hours, calibration with nitrogen (5.0 or higher N2 recommended)
shlf1314Zero point drift:< 2 % of measuring range per zero point calibration interval
Sensitivity drift:None
shlf1314Linearity deviation:< 2 % of measuring range
Temperature drifts:< 2 % of measuring range per 10K temperature change
shlf1314Pressure influence:1 % change of measuring value for 1 % sample pressure change.Ambient pressure changes measured and compensated
shlf1314零点校准:24H,用纯N2校准(推荐用5.0或更高的N2)
shlf1314零点漂移:< 2 %(小于每个零点的校准区间量程< 2 %)
灵敏度漂移:无
线性偏差:< 2 %(测量量程的2 %);
温度漂移:每10K的温度变化内漂移量<量程2%
shlf1314压力影响:样气压力变化1%,测量值变化1%,环境压力变化可被监测并得以补偿。
shlf1314Electrical connectors 电连接器
Digital interface:shlf13149-pole D-connector for RS-232 Analyzer is connected to an external computer via RS-232C cable. The external computer controls Gasmet.
Power connection:Standard plug CEE-22
数字接口:RS-232 D型9孔,分析仪通过 RS-232C 连接到外部计算机,由外部计算机控制傅立叶分析仪。
shlf1314电源连接:CEE-22标准插头
Gas inlet and outlet conditions 气体进出口条件
Gas temperature:Non-condensing, the sample gas temperature should be the same as the sample cell temperature
Flow rate:120 - 600 liters per hour
shlf1314Gas filtration:Filtration of particulates (2 μm) required
Sample gas pressure:Ambient
Sample pump:External, not included
shlf1314气体温度:无凝结,样气温度必须和样品室的温度相同
shlf1314流速:120-600L/h
shlf1314气体过滤器:过滤精度2 μm
样气压力:大气压
样气泵:外置,不包含
shlf1314Electronics 电气
A/D converter:Dynamic range 95 dB
Signal processor:32-bit floating point DSP 120 MFLOPS
Computer:External, not included
shlf1314A/D转换器:动态范围:95 dB
信号处理器:32 bit浮点 DSP 120 MFLOPS
计算机:外置,不包含
Analysis software (for external PC)
shlf1314Operating system:Windows 7 (32-bit)
shlf1314Analysis software:Calcmet for Windows
shlf1314分析软件(外部计算机)
shlf1314操作系统:Windows 7 (32位)
shlf1314分析软件:基于Windows的Calcmet
shlf1314Options 选项
Pressure measurement:Inside sample cell
External PC:
Gasmet PC for control and analysis with optional alarm relay boards and analog inputs or outputs
Sample cell gaskets:Kalrez®
压力测量:在样品室内部
shlf1314外置计算机:Gasmet工业电脑配备警报继电器板和模拟信号输入输出接口,用于系统控制和信号分析。
shlf1314Enclosure 附件
Sample cell material:Aluminium
Dimensions (mm):482 * 196 * 450
Weight:17 kg
CE label:According to EMI guideline 89/336/EC
样品室材质:铝
尺寸(mm):482 * 196 * 450
重量:17KG
shlf1314CE标签:据 EMI 89/336/ec 号指令